DTS
-
96
The electric waterproof membrane impervious meter is a new type of watertight instrument for testing petroleum asphalt linoleum and elastic full asphalt waterproofing membrane developed by our company.
GB328.3
-
89
The required test requirements. And suitable
GB18242
-
2000
,
GB18243
-
2000
Standard requirements. The instrument is pressurized by an electric air pump and can be used by relevant departments such as scientific research, construction, production, design and teaching.
This product adopts automatic control of gas and pressure, automatically maintains pressure and digital display timing, and provides very convenient equipment for testing, testing and researching the impermeability of linoleum and waterproof membrane. The test pressure is not greater than
0.6Mpa
In this case, the instrument can be used for the determination of other materials for gas permeability or water permeability. The following describes in detail how to use the electric waterproof membrane impervious meter.
How to use the electric waterproof membrane impervious meter
l
Adjust the pressure gauge according to the test requirements of the test piece before use.
The adjustment method is: there is an adjustment knob in the middle of the glass of the pressure gauge, and the upper and lower limits are adjusted, and a flat-blade screwdriver is needed. Screw the upper limit pointer to the position specified by the test, and the lower limit pointer is screwed to the lower limit.
0.05Mpa
The position so that the pressure is automatically stopped when the pressure reaches the demand (upper limit value) during work. When the leakage or permeable water causes the pressure to drop to a certain value (lower limit value), the air pump automatically starts the supplementary pressure. For example: test pressure requirements
0.3Mpa
, in the middle of the case, the upper limit pointer is dialed to the dial
0.3
Position, then use a screwdriver to screw the middle lower limit pointer to
0.25
s position.
2
When using, first use the hex wrench to loosen the pressure ring of the three permeable trays, unscrew the cover of the water inlet, and then close the drain valve, and slowly inject the water from the water inlet to the container.
2/3
Unscrew the valve separately
"1" "2" "3"
(
0
Water is poured out from the water inlet hole in the middle of the drain valve, overflowing the permeable tray, and then tightening the water inlet cap.
3
Place the tested piece on the annular rubber ring of the permeable tray, place the test piece pressure plate on the tested part, and tighten the screw.
4
Set the time dialing according to the experimental requirements and set the time value.
5
Plug in the power plug and turn on the start switch. The air pump starts to inject pressure gas into the container, and the pressure indicator lights up.
6
When the pressure upper limit is reached, the pressure is automatically stopped. At this time, the constant pressure indicator light is on. After the test is completed according to the specified time, the water valve is opened, the water is released and the pressure is released, and then the pressure ring is released to take off the test piece, and the test is completed.
Product Image
Previous:As part of asphalt fatigue testing machine equipment? -Huaqiang Electronic Network
Next: UV-1100 spectrophotometric method for rapid measurement of peroxide content of oxides in edible oils - Master's thesis - Dissertation
The material of this product is PC+ABS. All condition of our product is 100% brand new. OEM and ODM are avaliable of our products for your need. We also can produce the goods according to your specific requirement.
Our products built with input/output overvoltage protection, input/output overcurrent protection, over temperature protection, over power protection and short circuit protection. You can send more details of this product, so that we can offer best service to you!
Lcd Adapter,Mini Size Lcd Adapter,Security Lcd Adapter,Waterproof Lcd Adapter
Shenzhen Waweis Technology Co., Ltd. , https://www.huaweishiadapter.com